Resonance R26M-PT Diagnostic Middle Ear Analyzer w / Multifrequency Probe
Double-tap and hold to zoom.
Mouse over the image to zoom.

Resonance R26M-PT Diagnostic Middle Ear Analyzer w / Multifrequency Probe

Product Code: R26M-PT

  • Clinical, multi-frequency tympanometer with ipsi & contra acoustic reflex and decay capabilities
    • 226Hz 678 Hz, 800 Hz and 1000 Hz probe tones (pressure range +200 daPa to -400 daPa)
    • Animation feature for children when using 1000 Hz probe tone
    • Ipsi and contra acoustic reflex and decay testing capabilities at 0.5 1, 2 & 4 kHz (intensity range 50-110 dBHL)
  • 7" LCD, color graphic display
  • Powered by mains AC/DC adapter (included) that plugs into standard wall outlet
  • MDS software (included) allows data transfer to PC via Bluetooth 2.1 or USB cable, data storage (locally or in Noah) and reporting/printing features (NOTE: built-in printer within r26m-PT is obsolete)
  • Three year warranty
Other available options:
  • Eustachian Tube Function (ETF) and Acoustic Reflex Latency test capabilities: item# r36m-PT
  • optional external printer: item# O1000094
Compatible Eartips for Tympanometry     

     

  • Optional PPRT-02 Stand-alone Printer (item O1000094
  • Thermal paper for PPRT-02 Stand-alone Printer (item 8509893)
  • Immittance Tips
    • IT008 IT Series Eartips, 8mm Green
    • IT010 IT Series Eartips, 10mm, Red
    • IT012 IT Series Eartips, 12mm, Blue
    • IT014 IT Series Eartips, 14mm, Green
    • IT016 IT Series Eartips, 16mm, Red
    • IT001 IT Series Eartip Kit in Plastic Box (25 of each eartip size)

Parameter Set Up:

  • press F1 key to access set up menu
  • use knob or the down arrow to select parameter setup function of interest, and press ENTER
  • use side arrow to select options and/or change options

Available set up options

  • LANGUAGE: changes end-user interface language
  • DATE & TIME: changes system date and time
  • CONTRA REFLEX: selects type of transducer used for contra stimulation
  • REFLEX UP-DOWN: changes display of patient response during testing
  • AUTORUN: if enabled, test will start automatically once seal obtained
  • TYMP PROTOCOL: 2 options available.  Adult and Children.  Children is the fastest way to perform tympanometry.
  • LEAKAGE TEST: parameters may be adjusted to set to appropriate sensitivity 
  • REFLEX THRESHOLD: adjusts theauto reflex recognition sensitivity
  • SINGLE GRAPH: if enabled, right and left tympanogram reports printed separately 

To Perform Tympanometry:

  • Use arrow key to select tympanometry from the  main menu, then press ENTER
  • use SIDE key to select ear
  • if Autorun function active (via setup  menu), tympanometric testing will automatically start when a proper seal is achieved; otherwise, press START
  • When test complete, use SIDE key to select the other ear and repeat the process

To Perform Acoustic Reflex Testing: 

Acoustic reflex testing may be performed three different ways:

1. Manual mode: test is performed with user-defined parameters and results interpreted by provider

2. Automatic mode: test performed using preset parameters and results interpreted by provider

3. Auto Threshold mode: device automatically performs tests using preset parameters and results interpreted automatically according to sensitivity set in setup menu.

To perform acoustic reflex testing via manual mode:

  • select manual mode via corresponding function key (see item 4 in Controls and Components)
  • Use SIDE key to select ear
  • press IPSI or CONTRA keys to set stimulus presentation
  • use FREQUENCY keys to adjust
  • Use SIGNAL button to manually find threshold 

To Perform Acoustic Reflex Decay: 

Use arrows to select test from main menu; adjust test parameters as needed on main test screen.

Front Panel Controls & Components

1. display screen

2. built-in thermal printer (obsolete)

3. knob for manual pressure adjustment & main menu navigation

4. F1 - F6 function keys: enables execution of function displayed on the bottom section of the screen

5. TEST: when in test mode, this key enables return to main menu

6. ENTER: key for accessing tests from main menu

7. ARROW KEYS: used to select function on main menu

8. SIGNAL: manual start of reflex test

9. FREQUENCY

10. IPSI

11. CONTRA

12. SIDE: (right or left ear)

13. START/STOP

14. PRINT 

Back Panel Components

15. CONTRA transducer jack

16. PROBE connection

17. PRESSURE connector

18. FOOT SWITCH (obsolete)

19. USB host port connection

20. USB slave port connection for PC communication and data transfer

21. ON/OFF

ETF for intact and perforated tympanic membrane

  • Select ETF test from main menu
  • If autorun function is active, the tests will start automatically; otherwise, press START
  • After first tympanogram completed, instruct patient to swallow
  • Press CONTINUE to perform second tympanometer
  • After second tympanogram completed, instruct patient to swallow.
  • Press CONTINUE

Non-acoustic 

Test with graphical representation similar to the Decay test. Test length can be either of 10 or 20 secs. Thetest is conducted without acoustic stimulation, therefore the stapedial reflex can be stimulated by a tactileor ocular contraction with light air blows. This test can be adopted to investigate subjects with PatulousEustachian Tube. To change the length of the test press F5. The test can be conducted at 0 daPa or at tymppeak (maximum mobility) performing a tympanogram by pressing F1 PEAK (Zero or Auto).

Quick tests

Three rapid execution tests can be selected from the test selection page: Quick check, Quick screening,Quick decay. All three tests include the execution of a tympanogram and subsequently tests for thedetection of the stapedial reflex performed at the pressure at which the peak has been detected duringthe tympanogram . For all tests the various execution parameters can be changed by pressing F6 > F5.In the Quick check the stimuli for the reflexes will be emitted at the intensity selected in the parameterswindow, in the Quick Screening a reflex threshold research will be carried out with successive stimulationswith increasing intensity, the Quick decay will be performed using the same method of the screening plusthe decay test. The “ Children” mode can be activated in Quick Check by pressing F5. This feature, designedfor use with children, displays a sequence of colored images on the screen that are meant to attract thechild’s attention while performing the examination. When the image sequence is running, the test will berun in background, giving the operator the necessary information.

Growth - DLI

Similar to the Decay test, once the stimulus is issued at a fixed intensity, a series of short increment stimuliat variable intensity can be sent to the subject in order to highlight a further contraction of the stapedius.All controls are similar to those seen in the previous test for reflex detection.

HF tympanometry

Similar in execution and controls to the 226 Hz Tympanometry; three separate graphic areas are available to perform three different tympanograms at different probetone frequencies: 678 Hz, 800 Hz and 10000 Hz. The three tympanograms can be run with pressuregeneration in automatic or in manual mode.For each tympanogram, at the end of the test, three curves will be displayed shown. In tympanometry with 1000 Hz tone, the CHILDREN mode is available, as described in the Quick check.

How do I add my logo to the report?

  • go to MENU
  • select TOOLS
  • REPORT OPTIONS
  • REPORT CUSTOMIZATION
  • click the BUSINESS INFO tab. 
  • right click on image place holder to LOAD logo   

Things to Know Before you Buy

The built-in printer feature is obsolete and no longer functional

At the end of 2021, the printer was discontinued and Resonance was unable to find a comparable replacement that would fit in the unit.  Rather than redesigning the r26M form factor, the built-in printer comportment remains on current units. 

The unit does come packaged with free MDS software that allows data transfer from the audiometer to your PC for data management and report generation. The transferred data can be printed off from your PC. This represents the most common way providers print data from this unit. Another option for printing data from the unit is to purchase a compatible external printer that connects directly to the unit.  

Three  versions of the r26m are available; make the most informed decision by reading this

Three versions of the r26m include the r26m-ME, r26m, and r26m-PT.  All three offer tympanometry using 226 Hz probe tone.  The r26m-PT also offers additional probe tone frequencies (678, 800 and 1000 Hz).  Beyond this, the three units mainly differ according to acoustic reflex/decay capabilities.  The r26m-ME is limited to ipsilateral acoustic reflex testing and decay (no contra capabilities).  The r26m and r26m-PT offer ipsi and contra acoustic reflex testing and decay.   

   

This unit is essentially the same as the r36m-PT minus a few special tests/features

The r26m-PT offers the same capabilities as the r36m-PT with the exception of Eustachian Tube Function (ETF) tests.  The r36m-PT offers ETF test features that come with prompts to assist the provider in instructing the patient.  NOTE: it is possible to manually perform ETF tests on the r26m-PT without the special feature and something to consider when making a decision between the two models. 

Technical support, calibration, and service is provided and managed through Oaktree Products

Responsive, efficient, and effective support is critical for busy providers to be able to use their equipment without delay. Equipment calibration is required annually. Equipment may need to be serviced over the years.  Oaktree Products has made arrangements to coordinate and provide support, calibration and service to allow providers to focus on their clinical practice.

NOTE:

  • see PRODUCT DOCUMENTS tab and click SOFTWARE MANUAL to download MDS User Manual
  • QUICK START and QUICK START NOAH instructions outlined below after INSTALLATION

Minimal Requirements for PC to work with MDS Software:

  • Windows XP
  • Windows Vista
  • Windows 7
  • Windows 8

Minimal Requirements for Noah to work with MDS Software:

  • Noah version 4.X.

Resonance MDS Software & License Key:

The Resonance MDS Software is included free of charge with this Resonance unit. The software allows creation and management of a patient database on your PC and the ability to transfer patient information and test results from your Resonance unit to your PC. As a software program, a license key will be required to permanently activate the MDS software. The license key can be found within the Resonance License PDF file that resides in the MDS Suite folder on the USB thumb drive packaged with your unit. Print or refer to this file when your unit generates a prompt for MDS License Key Activation

MDS License Key Activation: 

The MDS software file resides in the MDS folder on the USB thumb drive packaged with your unit. The software is located on the USB thumb drive packaged with the unit.  Go to the MDS Suite folder, click MDS Setup and follow installation instructions. Be sure to click "Create Shortcut to Desktop" when prompted to have the software icon appear on your desktop for easiest access. NOTE: MDS software will accommodate data from five patients without requiring a software license activation key. When trying to add data from the sixth patient, the following Licensure Key activation window will appear:

This means the demo version of the MDS software has expired and it will be necessary to enter the License Key associated with your specific unit.  The MDS License Key is listed in the PDF file called Resonance License located within the MDS folder on the USB thumb drive that came packaged with the unit.  If you are unable to find the PDF file, contact Oaktree Products with the make/model and serial number of your unit.  We will make arrangements for the vendor to issue the necessary MDS License Key. 

Cables for Connection with PC:

The Resonance device can be connected to a PC using a USB cable (included)

INSTALLATION: 

  • Insert USB drive to PC
  • Double click file "MRS Resonance Setup.exe" to begin install
    • select YES if prompted with "Do you want to allow the following program from an unknown publisher make changes to your computer?"
  • Select language for installation from provided choices
  • At the Welcome screen, to continue, select NEXT
  • When prompted "Would you like to install this device software?", check the box next to "Always trust software from MRS Srl", then click INSTALL
  • When installation complete, select either "Run MRS Resonance Now" to run the program immediately, or click FINISH to complete installation and exit the program
  • The icon of the program will now be present on the PC desktop and in the program bar called MRS Resonance

Quick Start Instructions:

  • launch program
  • at the authentication window, "admin" will populate as UserName
    • type "a" (lower case letter a) as the password (password can be changed later)
    • check the box "remember me" to no longer have the authentication window populate when launching program
  • click SESSIONS (upper horizontal menu bar)

To Create a New Patient:

  • click NEW (from upper horizontal menu bar)
  • fill the the following MANDATORY FIELDS
    • First Name
    • Surname
    • Birth Date
    • Birth Place
    • Choose Gender
    • click PRIVACY ACCEPTED
  • click SAVE to create patient
  • click EDIT (to proceed with transferring data from Resonance device to your PC)

Transferring data from Resonance unit to PC:

  • after creating NEW patient, the icon "Acquire data from device" will be active
  • conduct test on the Resonance unit
  • connect Resonance device to the PC where MDS is running
  • click on "Acquire data from device"
  • an "Acquisition" or "Acquisizione" window will appear
    • select the Resonance device from pull down menu appearing under destination patient name
    • click DOWNLOAD
    • once the system indicates "Acquired Data", click SAVE & EXIT button to store data in database

View transferred test data:

Once data is transferred from a Resonance unit to a PC and saved, the test data will not reside in the '"Patient Historic Sessions" area located in lower left hand portion of the screen. To view the transferred data:

  • select SESSIONS tab located in the lower right hand portion of the screen under the image of the patient silhouette
  • in the TESTS IN SELECTION SESSION, click VIEW TEST DETAIL  

QUICK START NOAH:

  • launch MDS
  • click icon in the left upper corner of main window
  • select TOOLS > Noah > Register Module (this procedure executed one time only)
  • close Resonance software suite and run Noah
  • once Noah opened, a RESONANCE button will appear in the upper horizontal menu
  • to transfer data, follow instructions outlined above QUICK START INSTRUCTIONS, and select a patient in the Noah database
  • click RESONANCE button to run MDS
  • MDS will initiate; when the "Acquisition" or "Acquisizione" window appears, click DOWNLOAD to transfer data to Noah database